System Messages8 Troubleshooting
Message Recommendation
No Contrast e system was unable to ll the
syringe from the contrast container.
If the contrast container is empty,
replace it with a new container. If
there is contrast remaining in the
container, make sure that the white
slide clamp is open, that there are
no air bubbles in the spike, that the
contrast tubing is secured in the
contrast sensor, and that the contrast
sensor is clean.
Non-system disk or disk error e system has experienced an
internal fault that prevents it from
starting up normally. Contact a
service representative.
No X-ray Interface e expected communication with
a remote X-ray imaging system has
failed.
NVRAM Failure e system failed a self test during
startup.
Pressure Limit e system is adjusting the ow rate
so that the injection pressure remains
below the congured pressure limit.
Pressure Limit Exceeded Make sure that the stopcock is open
to allow injection and that the patient
catheter is not kinked. If appropriate,
increase the pressure limit, reduce
the ow rate, or increase the rise
time.
PURGE is Incomplete e syringe ram has reached its
maximum forward position before
the purge could be completed. Try
pressing START REFILL to initiate a
rell procedure, then run the purge
again. It may be necessary to replace
the patient kit.
Ram Forward (or Ram Reverse) is is an informational message
indicating that the syringe ram
has reached the forward or reverse
position.
Ready to Inject
Manifold Valve Open
Press Standby to Release
e injector will not activate if the
manifold valve is open. Press the
Standby button to allow system
pressure to relieve back pressure on
the manifold.
Release Trigger for New Injection Release the contrast (C) button on
the hand controller to begin a new
injection.