System Messages
78 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
8 Troubleshooting
Message Recommendation
Remove tubing from, and clean
contrast sensor
e system displays this alert when
it detects contrast during startup
calibration. If there is contrast tubing
in the sensor, remove it to continue
the setup procedure.
Reset X-ray interface for new
injection
Release the trigger on the connected
X-ray imaging system before the next
injection.
Reverse Limit Reached e syringe ram has reached its
rearmost position.
Saline Pump Malfunction (OS) e saline pump has experienced an
internal malfunction.
Saline Pump Malfunction (OT) e saline tubing may be obstructed
or pinched. Make sure the rear
tubing guide is securing the tubing
but not restricting ow of saline,
and that the front tubing guide is
positioned all the way up.
Standby! e Standby button has been
pressed, which has suspended all
system functionality. To resume
operation, press the Standby button
and then press OK on the control
panel.
Start Injection is is an informational message
displayed when an injection is
initiated from a connected X-ray
imaging system.
Syringe Not Detected e system has not detected a syringe
in the chamber. When prompted
during setup, insert a syringe into the
syringe chamber and fully close the
chamber door.
is message can also appear in error
if the syringe valve sensor needs to
be cleaned.
Syringe Valve Closed e syringe valve may be stuck.
Perform a short rell to try and
correct the problem.
If this alert occurs during setup,
the syringe may be faulty and may
need to be replaced. is alert can
also occur if the syringe valve sensor
needs to be cleaned.