ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 79
System Messages8 Troubleshooting
Message Recommendation
Syringe Valve Open e system displays this alert when
it detects a syringe already in the
chamber during startup calibration.
If there is a syringe in the chamber,
remove the syringe to continue the
setup procedure.
If there is not a syringe in the
chamber, the syringe valve sensor
may need cleaning or the sensor may
be blocked.
ere is a 10 second limit for Flush e system automatically stops a
saline ush aer 10 consecutive
seconds as a safety feature.
ere is a 10 second limit for Purge e system automatically stops a
purge aer 10 consecutive seconds as
a safety feature.
ere is a 20 minute limit for KVO e system automatically stops the
KVO function aer 20 consecutive
minutes as a safety feature.
Values Out of Range Try cycling power to the system.