130 Accessories and options
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
There are two types of FOT distinguished in fabian HFO:
FOT-HFO: The forced oscillation is created by the HFO module. The measurement does not
change the ventilation mode as it is available only in HFOV.
FOT-conventional: The forced oscillation is created by the exhalation membrane. During the
measurement, ventilation mode is changed to CPAP.
Ventilation
Mode
FOT
Type
HFO
FOT-HFO
CPAP
FOT-
Conventional
IPPV
PSV
SIMV
SIMV+PSV
SIPPV
FOT-HFO and FOT-conventional shares:
The same layout
The same operation steps
FOT-HFO and FOT-conventional differ in:
Parameter names
Parameter ranges
Forced signal creation
The following sensors and ventilation additives can be used together with FOT:
Volume Guarantee (VG)
Volume Limit (VL)
SpO
2
sensor
Predictive Intelligent Control of Oxygenation (PRICO)
The following option is disabled during the FOT recruitment / derecruitment maneuver:
Lung recruitment at HFOV