144 Sterilization / cleaning / disinfection
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
Neutralization
Wash
02:00
Warm tap water
Neodisher Z 1.0 ml per Liter
Rinse 1
01:00
Warm tap water
N/A
Dry Time
7:00
90°C / 194°F
(Set Point)
N/A
5. After the drying process package immediately for sterilization in packaging suitable for vapor
sterilization. Ensure that packaging stays dry before autoclaving.
Note: Articles must be individually single pouched in a qualified pouch (e.g. Cardinal Health self-
sealing pouch CAT# 92713). Each of the non-critical devices in-scope are thermostable and
have been validated for steam sterilization.
14.2.4.2 Sterilization options
The recommended sterilization method is steam sterilization by autoclave. The following validated
cycles are suitable for sterilization:
Option A:
1. Set the autoclave with the following parameters
4 preconditioning pulses
Steam cycle at 134°C / 273°F for 3 minutes
Dry cycle for 30 minutes
2. Individually single pouch the exhalation membrane and membrane holder separately.
3. Autoclave the exhalation membrane and membrane holder using the cycle parameters in Step 1
above.
Option B:
1. Set the autoclave with the following parameters
4 preconditioning pulses
Steam cycle at 134°C / 273°F for 4 minutes
Dry cycle for 30 minutes
2. Individually single pouch the exhalation membrane and membrane holder separately.
3. Autoclave the exhalation membrane and membrane holder using the cycle parameters in Step 1
above.
Option C:
1. Set the autoclave with the following parameters:
4 preconditioning pulses
Steam cycle at 134°C / 273°F for 5 minutes
Dry cycle for 30 minutes
2. Individually single pouch the exhalation membrane and membrane holder separately.
3. Autoclave the exhalation membrane and membrane holder using the cycle parameters in Step 1
above.
14.2.4.3 Duration of use
The exhalation membrane and membrane holder must be replaced after 30 sterilization cycles and also in
case of any detected defect that is affecting the accessory performance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals