146 Setting ranges and parameters
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
Mode
Parameter
SIMV
SIMV+PSV
Neonatal
Pediatric
Neonatal
Pediatric
min
max
min
max
min
max
min
max
Iflow [Lpm]
1
32
1
32
1
32
1
32
Eflow [Lpm]
1
32
1
32
1
32
1
32
Rise Time [sec]
0.1
2
0.3
2
0.1
2
0.3
2
Itime [sec]
0.1
2
0.3
2
0.1
2
0.3
2
Etime [sec]
0.5
30
0.5
30
0.5
30
0.5
30
Frequency (Rate) [1 / min]
2
100
2
100
2
100
2
100
PEEP [mbar]
0
30
0
30
0
30
0
30
Pinsp (backup)[mbar]
4
80
4
80
4
80
4
80
Ppsv [mbar]
2
80
2
80
Trigger
1
10
1
10
1
10
1
10
Vlimit [mL]
1
150
10
500
1
150
10
500
Vguarant [mL]
0.8
60
10
300
0.8
60
10
300
Pmax [mbar]
4
80
4
80
4
80
4
80
O
2
[%]
21
100
21
100
21
100
21
100
O
2
Flush [%]
23
100
23
100
23
100
23
100
Flush Time [sec]
0
120
0
120
0
120
0
120
Man. Breath Time [sec]
2
30
2
30
2
30
2
30
Termination criteria PSV [%]
1
85
1
85