Guide to volume guarantee 151
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
Step 3
Set in the Pmax area, the maximal pressure value to which the VG function can regulate.
Triggered and non-triggered breaths are independently supported based on lung compliance.
Normally, a triggered breath has a lower Pinsp than a non-triggered one due to the effort provided by
the patient.