Configurations menu 65
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
#
Parameter
Description
1.
Automatic O
2
Calibration
21 Vol% / 21 and 100 Vol.%. In cases where oxygen supply source
does NOT provide 100% oxygen, this should be set to 21% only to
avoid calibration at false value.
2.
NIV Leak
Compensation
Set either Off, Low, Middle or High level of leak compensation (0; up
to 10, 20, 30%) in nCPAP and DuoPAP modes.
3.
Unit for Pressure
Selection of Pressure unit (mbar / cmH2O).
4.
NIV Tube Set
Selection of the nCPAP system used (Infant Flow (Infant Flow or
Inspire), Infant Flow LP, Medijet
®
).
5.
Maximum time
for Manual breath
Maximum Inspiration time for Manual breath (2 to 30 seconds).
6.
Alarm Delay
After visual reset of alarm, the reactivation of alarm is delayed for
the set time period. This avoids an immediate re-alarm in case
Operational parameters have not yet stabilized.
7.
Manual Breath
HFO
ON / OFF. The Manual Breath in HFO mode can be used to perform
Lung Recruitment Maneuver (sustained lung inflation) if clinically
accepted by internal guideline. Length and Pressure level are
adjustable.
8.
Trigger
(conventional)
Flow Trigger / Volume Trigger / Pressure Trigger. Trigger mode
Configuration for assisted breathing
9.
E-flow
E-flow parameter setting
1
2
3
4
5
6
7
8
9
10
11
12
13
14

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals