15.2.2.3 Sterilization options
The recommended sterilization method is steam sterilization by autoclave. The
following validated cycles are suitable for sterilization:
Option A:
1. Set the autoclave with the following parameters
1.1. 4 preconditioning pulses
1.2. Steam cycle at 13C / 273°F for 3 minutes
1.3. Dry cycle for 30 minutes
2. Individually single pouch the cassette articles
3. Autoclave the cassette articles using the cycle parameters in Step 1
Option B:
1. Set the autoclave with the following parameters
1.1. 4 preconditioning pulses
1.2. Steam cycle at 13C / 273°F for 4 minutes
1.3. Dry cycle for 30 minutes
2. Individually single pouch the cassette articles
3. Autoclave the cassette articles using the cycle parameters in Step 1
Option C:
1. Set the autoclave with the following parameters:
1.1. 4 preconditioning pulses
1.2. Steam cycle at 13C / 273°F for 5 minutes
1.3. Dry cycle for 30 minutes
2. Individually single pouch the cassette articles
3. Autoclave the cassette articles using the cycle parameters in Step 1