115
To stop this test, press the back button .
4. Save the result
Click the save button to save the test results.
Required fields are as follows: Name and serial number, then click OK button to save.
Click the back button to return to the test type list interface to exit without saving the results.
To restart the test, click the reset button to return to the ready to start test interface.
5.3.3 Aneroid barometer
1. Prepare
Select “aneroid barometer” in the test type list to enter to the quick test process of the aneroid
barometer.
2. Connection
Connect to the calibrator as shown in Figure 5-9.
3. Start the test
Click the start button to select the automatic execution to enter the execution parameter
setting interface. Fill in the set point list, the number of cycles, the stroke mode, the number of
readings, the reading interval, and the dwell time, and then click the OK button to start the test.
During the test, manually input the reading and correction value of the aneroid barometer.
Click the next point button to skip the current set point and advance to the next set point.