control.
Selecting manual execution requires the test point to be entered as the test progresses.
To stop this test, click the back button .
3. Save the result
Click the save button to save the test result.
The following items will need to be entered: Name and serial number. Then click the check mark in
the lower right corner to complete the savethe results.
To exit without saving press the back button .
To redo the test click the reset button .
5.1.2 Pressure transmitter
Select the “pressure transmitter” icon in the test type list to start the quick.
1. Connection
Refer to Section 3.6.2 for connection.
2. Start the test
Click the zeroing button to reset the pressure control module.
Click the start button to select the automatic execution to enter the execution parameter
setting interface. Fill in the set point list, the number of cycles, the stroke mode, the number of
readings, the reading interval, and the dwell time, and then click the OK button to start the