3.6 Typical applications
3.6.1 Pressure gauge
1. Prepare
Open the pressure output area of main operation interface: High-pressure or low-pressure range
(depend on the range of calibrated instrument).
2. Connection
Connect the gauge as shown in Figure 3-12.
When the calibrated instrument is a differential pressure gauge, connect the REF port properly.
For the connection mode, see Figure 3-2 Differential pressure output.
3. Manually set or use the step function (see Section 3.2.1 "Auto step and manual step of pressure
output") to output the pressure of each calibration point.