116
Selecting the manual execution requires manual input of each set point and readings.
To stop the test, press the back button .
4. Save the result
Click the save button to save the test results.
Required fields are as follows: Name and serial number, then click OK button to save.
Click the back button to return to the test type list interface to exit without saving the results.
To restart the test, click the reset button to return to the ready to start test interface.
5.3.4 Barometric altimeter
1. Prepare
Select “barometric altimeter” in the test type list to enter to the quick test process of the barometric
altimeter.
2. Connection
Connect to the calibrators shown in Figure 5-10.
3. Start the test
Click the start button to select the automatic execution to enter the execution parameter setting
interface. Fill in the set point list, the number of cycles, the stroke mode, the number of readings,
the reading interval, and the dwell time, and then click the OK button to start the test. During
the test manually input the reading of the barometric altimeter.