109
gauge.
When starting the electric contact test in the task setting, it will test the on/off value, and during
the test, it will switch to the data collection interface to collect the switch on and off values.
Enable or disable loop power as required for the input.
For Enable or disable loop power as required for the output.5. Task end
For the contents, refer to Section 5.2.1 Task end for calibration of the analog pressure gauge.
5.2.9 Valve opening gauge
1. Connection
Connect the calibrator and the valve opening gauge as shown in Figure 5-5.
2. New task
Select the “valve opening gauge” icon in the test type list, and input the information of the
calibrated valve opening gauge.