61
The temperature can be input within the range of (0~100) °C.
8. Set point limit
Input a range smaller than the calibrator range as the upper and lower limits of pressure control, to
ensure that the pressure output to the calibrated device does not exceed its range, thereby
protecting the calibrated device.
If disabled, this function is invalid. The setting will be stored by the calibrator and still effective after
reboot.
If enabled, the target value of pressure control of the calibrator is restricted by this condition.
When exceeding this condition, the calibrator will give a prompt that the set point limit is enabled
and the designated pressure cannot be outputted.
4.2 Communication
The calibrator supports three communication modes: Ethernet, Wi-Fi and Bluetooth (BLE).
4.2.1 Ethernet
The calibrator is connected via network cables to computers for communication.
Select the acquisition mode of device address.
Table 4-1 Network connecting