After the power supply configuration is selected, you will enter the search interface. Start
searching HART devices from the address "0". If successful the connection will be made and the
HART device will be displayed. If no connection is established the calibrator will continue to search
from the address "1" until "15". After the search is complete, the calibrator will list all the HART
devices found, and can support up to 15 HART devices at the same time.
During the search process, you can press the key to stop searching and return to the power
configuration interface.
After searching is finished, you can press the key to conduct a new search.
After searching is finished, if any HART device is online, press the / key to select it, then
press the key to confirm establishing connection with this device; press the key and
select "Setup" to read main information of the selected device.
2. Online/offline
After connection is established, HART will be displayed as electrical measurement item on the
main interface.
Under the main interface, when electrical measurement is switched to other measurement mode
(e.g. current measurement), it will exit HART connection. Meanwhile, the HART measurement icon
in the status bar will change to a corresponding icon (e.g. current measurement).
Under the main interface, if HART communication fails, it will automatically search new devices