User Manual MIC 900 / MIC 1100 / MIC 3000
5.1.3 Starting a program with batch number
If the Batch no function is enabled, the controller will ask you to enter a batch number description.
Entering a batch number
Press the top line in the batch no. box. An on-screen keypad will appear and you can enter the batch
description. Then press “Enter” to confirm your entry and to apply the description to the current
program.
A batch no. has to be entered when this function was enabled. Without a batch number the
program cannot be started.
After you have entered and confirmed the batch number you can choose between: „Start with time “
or „Instant start “.
5.1.4 Sensor log on or off (Sensor registration)
This option is only available when at least one sensor exists, which can be logged on at the
start of a program.
Logging on at the start of a program
Every time a program is started, a window is shown with
sensors that can be activated. Click on the relevant box
in front of the sensor name to activated the sensor („Box
is filled“) or to deactivated it („Box is empty“).
Log on manually
This window for sensor registration is not only shown at
the start of a program, it can also be called up by
pressing „Opt. control loops“ (at any time) and the
window „Sensor registration“ will appear.
Sensor display
The „de-activated“ sensors are „highlighted“ (grey background)
[-] – Error „No feedback“
If no sensor has been activated for a measured variable, the sensor error „No feedback”
is shown for this measured variable.