User Manual MIC 900 / MIC 1100 / MIC 3000
10 List of errors (possible problems)
Set up a program and assign at
least one process to the first step
(See „
Programming mode“, item 0, page
21).
Enter a waiting time („Start“
button is flashing)
Press the „Start“ button.
Program is terminated
immediately after the start
Possible defect of the chamber,
core or humidity sensor, display
shows „EEE“ for sensor break or
„---“ for sensor short circuit
Replace the broken sensor or call
your service technician.
After it has been replaced the
temperatures have to be checked
and the controller may have to be
re-calibrated by the service
technician.
Program runs quickly through
all the steps and then ends
- the pre-set core temperature
values was reached
- FC value was reached
Check the pre-set values in this
program.
- nominal humidity value has
been reached (humidity min.),
- Nominal de-humidification value
has been reached (humidity
max.)
If it has been configured that way ,
ask your service technician
Nominal core value is higher than
the pre-set nominal chamber
value (negative core shut-down,
core temperature is below the
nominal chamber temperature)
See See „Negative core shut-down
(Showering/cooling)“, item 6.1.1,
page 32
Program does not switch not
the next step
Single step control is active
Manually switch into the next step,
if you want to go onto the next step
or increase the nominal values if
you want to continue with the
current process.
(See „Single step control“, item
6.6, page 34)
Program steps are repeated
Step repetition sequence is
active
Push the „Stop“ button
(See „Setting up repetitive steps“,
item 4.3.5, page 24)
A program that has just finished
automatically switches into the
next program
Confirm with „Stop“
(See „Linking programs“, item 4.4,
page 26)
Controller cannot be started
Disconnect the controller from the
power supply for at least one
minute, then reconnect it and try
operating it again.
CAN BUS error
CAN BUS will be blocked,
Error message: „Error at CAN-
BUS (No.: 0x1007). Modules
cannot be used“
The message will remain on the
display until the controller is
rebooted.
Modules of the same type have
the same CAN addresses
(see switch settings (DIP switch)
for CAN modules).
Change the module addresses (DIP
switch).