EasyManua.ls Logo

aditec MIC 3000 - Page 43

Default Icon
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual MIC 900 / MIC 1100 / MIC 3000
43
Issue 17.08.2021_00
8.7.1.1 Changing user names, access rights, global administrator settings
Administrator password:
You will need to enter a password in order to configure or change the following user settings. This
password can be individually changed for each installation. Factory setting is „7890“.
For security reasons, administrator passwords should be changed.
Settings:
Administrators
Enabling or disabling administrators (ON/OFF)
Automatic log-off
Settings for operating mode
- Time interval is pre-programmed: (Range of entries: 0 to 254 sec.)
If none of the keypad buttons are pressed, the controller will automatically log-off the current
user after this time interval has elapsed. This means the user will have to re-register with the
controller if she/he would like to make further changes.(e.g. to switch into the next step, to make
changes to the nominal values etc.)
- No interval is pre-programmed:
If you have not entered a time interval, the current user will not be logged off automatically. This
means that further tasks of the latest user can be performed until he manually logs off
him/herself.
Hint
Manual log-off in operating mode
Menu / Profiles/ „Admin.“ button
See Display of , page 44
Administrator password
2
1
Once a correct password was entered, the headings (except for the
PIN) are shown in green. The settings can now be changed.
Also see „Changing the user PIN“ Item 8.7.1.2, Page 44
Enter the administrator password in order to
enable the user settings.
1
2
3

Table of Contents

Related product manuals