EasyManuals Logo

Advanced Instruments OsmoTECH User Manual

Advanced Instruments OsmoTECH
67 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
59
OsmoTECH® Single-Sample Micro-Osmometer User Guide
Appendix C: Known limitations
Additives such as anticoagulants from collection tubes may contribute to the measured osmolality.
Particulate matter can cause premature crystallization, such as sample pre-freeze. The particulate matter may be
eliminated by centrifugation. It may also be eliminated in urine by filtration.
Samples with air bubbles do not freeze reliably and may yield inaccurate results.
Samples allowed to evaporate will yield higher osmolality values.
Non-aqueous solutions will not yield accurate results. The OsmoTECH is intended to test aqueous solutions only.
Highly viscous solutions may not freeze reliably and may yield inaccurate results.
High concentrations of ethanol, acetone, methanol, isopropanol, ethylene glycol, diethyl ether, paraldehyde,
trichloroethane, or propylene glycol may not freeze reliably and may yield inaccurate results.

Table of Contents

Other manuals for Advanced Instruments OsmoTECH

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoTECH and is the answer not in the manual?

Advanced Instruments OsmoTECH Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoTECH
CategoryMedical Equipment
LanguageEnglish

Related product manuals