Agilent 1260 Infinity DAD and MWD User Manual 121
Troubleshooting and Diagnostics
6
User Interfaces
User Interfaces
Depending on the used interface, the available tests and the screens/reports may vary.
Preferred tool should be the Agilent Diagnostic Software, see “Agil ent Lab A dvi sor
Software” on page 122.
Screenshots used within these procedures are based on the Agilent ChemStation.
In future, a user interface may not show the Diagnostics/Tests anymore. Then the Agilent
Diagnostic Software must be used instead.
The Agilent ChemStation may not include any maintenance/test functions.
Tabl e 1 8 Test Functions avaible vs. User Interface
Test Diagnostic Software ChemStation Instant Pilot G4208A
Selftest Yes Yes No
Filter Yes Yes No
Slit Yes Yes No
D/A Converter Yes Yes No
Tes t C hr om a to gr am Yes Ye s ( *) No
Wavelength Calibration Yes Yes Yes
Lamp Intensity Yes Yes Yes
Holmium Yes Yes Yes
Cell Yes Yes Yes
Dark Current Yes Yes Yes
* requires a command via command line