Agilent 1260 Infinity DAD and MWD User Manual 263
LAN Configuration
13
TCP/IP parameter configuration
TCP/IP parameter configuration
To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
IP address
Subnet Mask
Default Gateway
The TCP/IP parameters can be configured by the following methods:
by automatically requesting the parameters from a network-based BOOTP
Server (using the so-called Bootstrap Protocol)
by automatically requesting the parameters from a network-based DHCP
Server (using the so-called Dynamic Host Configuration Protocol). This
mode requires a LAN-onboard Module or a G1369C LAN Interface card, see
“Setup (DHCP)” on page 270
by manually setting the parameters using Telnet
by manually setting the parameters using the Instant Pilot (G4208A)
The LAN interface differentiates between several initialization modes. The
initialization mode (short form ‘init mode’) defines how to determine the
active TCP/IP parameters after power-on. The parameters may be derived
from a Bootp cycle, non-volatile memory or initialized with known default
values. The initialization mode is selected by the configuration switch, see
Table 29 on page 265.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals