82 Agilent 1260 Infinity DAD and MWD User Manual
4
Using the Detector
Special Settings of the Detector
Automated Run Recovery in case of temporary communication
failures
Tabl e 1 2 Automated Run Recovery in case of temporary communication failures
Situation Reaction On ChemStation
All OK • Run in progress - Data Analysis
• Run / Rawdata
• Elapsed run time is running
• Spectra counter is running
• Data are stored on PC and on card
LAN breaks • Run in progress - Data Analysis
• Run / Rawdata
• Error Power Fail
• Elapsed run time stops
• Spectra counter stops
• Data continues to be stored on card
LAN recovers • Run in progress - Data Analysis
• Run / Rawdata
• Error Power Fail cleared
• Elapsed run time continues at actual time
• Data continues to be stored on PC and on card
• ChemStation tries already to add missing data (depends on the
data load).