188 Agilent 1260 Infinity DAD and MWD User Manual
9
Maintenance
Replacing Capillaries on a Semi-Micro and Micro Flow Cell
Replacing Capillaries on a Semi-Micro and Micro Flow Cell
When If the capillary is blocked
Tools required Description
Wrench, 1/4 inch
for capillary connections
Wrench, 4 mm
(for capillary connections)
Screwdriver, Pozidriv #1 PT3
Parts required Description
For parts see “Semi-Micro Flow Cell Parts” on page 212 or “Micro Flow Cell” on page 214.
Preparations Turn th e lamp(s) off.
Remove the front cover.
Remove the flow cell, “Exchanging a Flow Cellon page 172.
NOTE
All descriptions in this procedure are based on the default orientation of the cell (as it is
manufactured). The heat exchanger/capillary and the cell body can be fixed mirror
symmetrically to have both capillaries routed to the bottom or to the top (depending on the
routing of the capillaries to the column).
NOTE
The fittings at the flow cell body are special types for low dead volumes and not
compatible with other fittings.
When retightening the fittings, make sure that they are carefully tightened (handtight plus
1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals