134 Agilent 1260 Infinity DAD and MWD User Manual
7
Error Information
Detector Error Messages
Detector Error Messages
These errors are detector specific.
Visible Lamp Current
The visible lamp current is missing.
The processor continually monitors the lamp current during operation. If the
current falls below the lower current limit, the error message is generated.
Visible Lamp Voltage
The visible lamp voltage is missing.
The processor continually monitors the voltage across the lamp during
operation. If the lamp voltage falls below the lower limit, the error message is
generated.
Probable cause Suggested actions
1
Lamp disconnected.
Ensure the visible lamp connector is seated firmly.
2
Defective visible lamp.
Exchange the visible lamp.
3
Defective connector or cable.
Please contact your Agilent service representative.
4
Defective power supply.
Please contact your Agilent service representative.
Probable cause Suggested actions
1
Defective connector or cable.
Please contact your Agilent service representative.
2
Defective power supply.
Please contact your Agilent service representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals