EasyManuals Logo

Agilent Technologies 1260 Infinity G1315D User Manual

Agilent Technologies 1260 Infinity G1315D
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #197 background imageLoading...
Page #197 background image
Agilent 1260 Infinity DAD and MWD User Manual 197
Maintenance
9
Cleaning or Exchanging the Holmium Oxide Filter
Cleaning or Exchanging the Holmium Oxide Filter
When If holmium oxide filter is contaminated
Tools required Description
Screwdriver, Pozidriv #1 PT3
Screwdriver, flat blade
Wrench, 1/4 inch
for capillary connections
Pair of tweezers
Parts required # p/n Description
1 79880-22711 Holmium oxide filter
Preparations Turn th e lamp(s) off.
Remove the front cover.
Remove the flow cell, see “Exchanging a Flow Cell” on page 172.
NOTE
See also “Declaration of Conformity for HOX2 Filter” on page 301.
The glass tends to build a film on its surface even under normal environmental conditions.
This is a phenomenon, which can be found also on the surface of several other glasses
and has something to do with the composition of the glass. There is no indication, that the
film has an influence on the measurement. Even in the case of a thick film, which scatters
the light remarkably, no shift of the peak positions is to be expected. A slight change in the
absorbance might be possible. Other components within the light path (lenses,
windows, ...) are also changing their behavior over the time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals