Nano Flow Cell - Replacing or Cleaning
Figure 60 Wrench plus Torque Adapter
NOTE
With the instrument accessory kit comes a 4-mm wrench and with the Sealing Kit a
special adapter. Both together work as a torque wrench with pre-defined torque (maximum
allowed torque for the cell fittings is 0.7 Nm). It can be used to tight the capillary fittings at
the flow cell body. The wrench has to be plugged into the adapter as shown in Figure 60
on page 196.
:UHQFK
$GDSWHU
'2127SUHVVGRZQ
PRUHWKDQVKRZQKHUH
PD[
1P