EasyManuals Logo

Agilent Technologies 1260 Infinity G1315D User Manual

Agilent Technologies 1260 Infinity G1315D
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
24 Agilent 1260 Infinity DAD and MWD User Manual
2
Site Requirements and Specifications
Performance Specifications
Performance Specifications
Specifications
Performance Specifications G1315C
Tabl e 3 Performance Specifications G1315C
Type Specification Comments
Detection type 1024-element photodiode array
Light source Deuterium and tungsten lamps The UV-lamp is equipped with
RFID tag that holds lamp typical
information.
Data rate up to 80 Hz
Wavelength range 190 – 950 nm
Short term noise
(ASTM) Single and
Multi-Wavelength
< ± 0.7·10
-5
AU at 254 and 750 nm
see "Specification Conditions"
below
Drift
< 0.9·10
-3
AU/h at 254 nm
see "Specification Conditions"
below
Linear absorbance
range
> 2 AU (5 %) at 265 nm see "Specification Conditions"
below
Wavelength
accuracy
± 1 nm Self-calibration with deuterium
lines, verification with holmium
oxide filter
Wavelength
bunching
1 – 400 nm Programmable in steps of 1 nm
Slit width 1, 2, 4 , 8, 16 nm Programmable slit
Diode width < 1 nm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals