Agilent 1260 Infinity DAD and MWD User Manual 139
Error Information
7
Detector Error Messages
Illegal Temperature Value from Sensor at Air Inlet
Error ID: 1072
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100 centigrade.
As a result the temperature control is switched off.
Wavelength Recalibration Lost
Error ID: 1037
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If
no data is available in the spectrometer ROM, the error message is generated.
Probable cause Suggested actions
1
The temperature sensor is defect.
• Replace the cable to the main board.
• Please contact your Agilent service
representative.
2
Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within
the allowed range.
Probable cause Suggested actions
1
The detector is new.
Recalibrate the detector.
2
The detector has been repaired.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals