Test Functions
Self-test
Self-test
The DAD self-test (see Figure 46 on page 144) runs a series of individual tests,
and evaluates the results automatically. The following tests are run:
• Filter Test
• Slit Test
• Dark Current Test
• Intensity Test
• Wavelength Calibration Test
• Holmium Test
• Spectral Flatness Test
• ASTM Noise Test (optional)
Figure 46 Self-test
For details refer to the individual tests on the following pages.