136 Agilent 1260 Infinity DAD and MWD User Manual
7
Error Information
Detector Error Messages
UV Lamp Voltage
Error ID: 7451
The UV lamp anode voltage is missing.
The processor continually monitors the anode voltage across the lamp during
operation. If the anode voltage falls below the lower limit, the error message is
generated.
UV Ignition Failed
Error ID: 7452
The UV lamp failed to ignite.
The processor monitors the UV lamp current during the ignition cycle. If the
lamp current does not rise above the lower limit within 2 – 5 seconds, the
error message is generated.
Probable cause Suggested actions
1
Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
2
Defective detector main board.
Please contact your Agilent service representative.
3
Defective power supply.
Please contact your Agilent service representative.
Probable cause Suggested actions
1
Lamp too hot. Hot gas discharge lamps
may not ignite as easily as cold lamps.
Switch off the lamp and allow it to cool down for at
least 15 minutes.
2
Lamp disconnected.
Ensure the lamp is connected.
3
Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
4
Defective detector main board.
Please contact your Agilent service representative.
5
Defective power supply.
Please contact your Agilent service representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals