192 Agilent 1260 Infinity DAD and MWD User Manual
9
Maintenance
Nano Flow Cell - Replacing or Cleaning
Nano Flow Cell - Replacing or Cleaning
When If parts are contaminated or leaky.
Tools required Description
Screwdriver, Pozidriv #1 PT3
Wrench, 1/4 inch
for capillary connections
Parts required Description
For parts identification refer to “Nano Flow Cells” on page 220 (80 nL and 500 nL).
Preparations Turn th e lamp(s) off.
Remove the front cover.
Remove the flow cell, see “Exchanging a Flow Cell” on page 172.
NOTE
For details refer to the technical note that comes with the nano-flow cell kit.
NOTE
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet windows
at the quartz block.