Agilent 1260 Infinity DAD and MWD User Manual 191
Maintenance
9
Replacing Capillaries on a Semi-Micro and Micro Flow Cell
9
Check for a centered holder vs. hole. If required adjust
with the holder screws.
Next Steps:
10
Reconnect the capillaries, see “Exchanging a Flow
Cell” on page 172.
11
Perform a leak test.
12
Insert the flow cell.
13
Replace the front cover.
14
Perform a “Wavelength Verification and Calibration” on
page 160 or a “Holmium Oxide Test” on page 151 to
check the correct positioning of the flow cell.