EasyManuals Logo

Agilent Technologies 1260 Infinity G1315D User Manual

Agilent Technologies 1260 Infinity G1315D
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
138 Agilent 1260 Infinity DAD and MWD User Manual
7
Error Information
Detector Error Messages
Holmium Oxide Test Failed
Illegal Temperature Value from Sensor on Main Board
Error ID: 1071
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable cause Suggested actions
1
Lamps switched off.
Ensure the lamps are switched on.
2
Defective or dirty flow cell.
Ensure the flow cell is inserted correctly, and is
free from contamination (cell windows, buffers
etc.).
3
Defective filter assembly.
Please contact your Agilent service
representative.
4
Defective achromat assembly.
Please contact your Agilent service
representative.
5
Defective PDA/optical unit.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Defective sensor or main board.
Please contact your Agilent service
representative.
2
Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within
the allowed range.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals