Error Information
Detector Error Messages
Holmium Oxide Test Failed
Illegal Temperature Value from Sensor on Main Board
Error ID: 1071
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable cause Suggested actions
1
Lamps switched off.
Ensure the lamps are switched on.
2
Defective or dirty flow cell.
Ensure the flow cell is inserted correctly, and is
free from contamination (cell windows, buffers
etc.).
3
Defective filter assembly.
Please contact your Agilent service
representative.
4
Defective achromat assembly.
Please contact your Agilent service
representative.
5
Defective PDA/optical unit.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Defective sensor or main board.
Please contact your Agilent service
representative.
2
Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within
the allowed range.