Agilent 1260 Infinity DAD and MWD User Manual 133
Error Information
7
General Error Messages
2 If it is desired that the system has only one leak drain tube, then it's
possible in increase the length of the small plastic tube which is connected
to port# 4 of the sampler switching valve. This tube can then be taken to
waste separately. The tube which normally serves as the detector cell outlet
tube can be used for this purpose.
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed, the fan is switched off, and the error message is generated.
Cover Violation
Error ID: 7461
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed while the lamps are on (or if an attempt is made to switch on
for example the lamps with the foam removed), the lamps are switched off,
and the error message is generated.
Probable cause Suggested actions
1
The top foam was removed during
operation.
Please contact your Agilent service representative.
2
Foam not activating the sensor.
Please contact your Agilent service representative.
3
Defective sensor or main board.
Please contact your Agilent service representative.
Probable cause Suggested actions
1
The top foam was removed during
operation.
Please contact your Agilent service representative.
2
Foam not activating the sensor.
Please contact your Agilent service representative.