Agilent 1260 Infinity DAD and MWD User Manual 147
Test Functions
8
Filter Test
Test Evaluation
Filter Test Failed
Test Failed
Holmium Oxide Maximum out of Limits
Probable cause Suggested actions
1
Filter assembly (lever and filter) not
installed.
Install the filter assembly.
2
Defective filter motor.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Holmium oxide filter not installed.
Install the holmium oxide filter.
2
Dirty or contaminated filter.
Exchange the holmium oxide filter.