Agilent 1260 Infinity DAD and MWD User Manual 189
Maintenance
9
Replacing Capillaries on a Semi-Micro and Micro Flow Cell
1
Identify the inlet and outlet capillaries.
2
After replacing the outlet capillary, fix it handtight first.
Then do a 1/4 turn with a 4-mm wrench.
3
To replace the inlet capillary, use a 4-mm wrench for the
fitting.
4
Unscrew the cell body from the heat exchanger and the
heat exchanger from the clamp unit.
,QOHWFDSLOODU\
2XWOHWFDSLOODU\