Agilent 1260 Infinity DAD and MWD User Manual 149
Test Functions
8
Intensity Test
Figure 49 Intensity Test
Figure 50 Intensity Test (Signal)
In case of low counts in one or more ranges, start the testing with the
comparison of values with flow cell vs. flow cell removed.
Contaminations of the cell windows and/or the lenses (there are 3 between
vis-lamp and flow cell), will reduce the light throughput.