Test Functions
8
ASTM Drift and Noise Test
Test Failed
Probable cause Suggested actions
1
Insufficient lamp warm-up time
Allow lamp to warm-up for at least 1 hour.
2
Old or non-Agilent lamp.
Exchange the lamp.