Agilent 1260 Infinity DAD and MWD User Manual 173
Maintenance
9
Exchanging a Flow Cell
1
Press the release buttons and remove the front cover to
gain access to the flow cell area.
2
Open the flow cell cover.
Depending on the system setup, the inlet capillary
might be routed directly from the module above or
below to the cell and not to the capillary holder.
3
Disconnect the flow cell inlet capillary (top) and the
waste tubing (bottom) from the unions.
4
Loosen the thumb screw and remove the flow cell outlet
capillary (bottom) with the union.