Agilent 1260 Infinity DAD and MWD User Manual 175
Maintenance
9
Exchanging a Flow Cell
7
Insert the flow cell capillaries into the union holder (top
is inlet, bottom is outlet). Tighten the thumb screw and
Reconnect the waste tubing (bottom) to the union.
To check for leaks, establish a flow and observe the
flow cell (outside of the cell compartment) and all
capillary connections.
Next Steps:
8
Perform a “Wavelength Verification and Calibration” on
page 160 or a “Holmium Oxide Test” on page 151 to
check the correct positioning of the flow cell.
9
Replace the front cover.