204 Agilent 1260 Infinity DAD and MWD User Manual
9
Maintenance
Replacing the Module’s Firmware
Module Specific Information
Tabl e 2 0 Module Specific Information (G1315C/D and G1365C/D)
G1315C DAD VL+ / G1365C
MWD
G1315D DAD / G1365D MWD
Initial firmware (main
and resident)
B.01.02 B.01.04
Compatibility with
1260/1290 Infinity
modules
When using the G1315C/D and G1365C/D in a system, all other
modules must have firmware revision A.06.xx or B.06.xx or above (main
and resident) from the same revision set (e.g. A.06.30/B.06.30).
Compatibility with
1100/1200 series
modules
When using the G1315C/D and G1365C/D in a system, all other
modules must have firmware revision A.06.xx or B.01.02 or above (main
and resident). Otherwise the communication will not work.
Compatibility with VSA
Optical
Introduced 08/2012. Firmware B.06.51, B.06.43 or B.06.26 or later
(depends on the used firmware set). Earlier revisions are not
compatible with the VSA Optical. These revisions are the required
versions for the new VSA Optical Unit and Main Boards.
Conversion to /
emulation of G1315B or
G1365B
Not possible due to different hardware and electronic platform.