Agilent 1260 Infinity DAD and MWD User Manual 31
Site Requirements and Specifications
2
Performance Specifications
Flow cells Standard: 13 µL volume, 10 mm cell
path length and 120 bar (1740 psi)
pressure maximum
Standard bio-inert: 13 µL volume,
10 mm cell path length and 120 bar
(1740 psi) pressure maximum
Semi-micro: 5 µL volume, 6 mm cell
path length and 120 bar (1740 psi)
pressure maximum
Micro: 2 µL volume, 3 mm cell path
length, 120 bar (1740 psi) pressure
maximum
Semi-nano: 500 nL volume, 10 mm
cell path length and 50 bar (725 psi)
pressure maximum
Nano: 80 nL volume, 6 mm cell path
length and 50 bar (725 psi) pressure
maximum
High pressure: 1.7 µL volume, 6 mm
cell path length and 400 bar
(5800 psi) pressure maximum
Prep SST: 3 mm cell path length and
120 bar (1740 psi) pressure maximum
Prep Quartz:0.3 mm cell path length
and 20 bar (290 psi) pressure
maximum
Prep Quartz: 0.06 mm cell path length
and 20 bar (290 psi) pressure
maximum
All flow cells are equipped with
RFID tags that hold cell typical
information.
pH range 1.0—9.5 (12.5 solvent
dependent with bio-inert version)
Time programmable Wavelength, polarity, peak width,
lamp bandwidth, autobalance,
wavelength range, threshold, spectra
storage mode
Tabl e 5 Performance Specifications G1365C
Type Specification Comments