Agilent 1260 Infinity DAD and MWD User Manual 53
Installing the Module
3
Flow Connections to the Detector
The installation of the detector is complete now.
5
If another Agilent module is positioned on top of the
detector, route the tubing assembly waste from the
accessory kit behind the capillary holder and connect the
top end to the other module’s waste outlet.
6
Connect the capillary from the column to the capillary
holder (top). Connect the PTFE waste tubing to the flow
cell outlet fitting (bottom) and the corrugated waste
tubing to the leak outlet.
7
Remove the flow cell and establish a flow and observe
for leaks.
8
Insert the flow cell, close the cover and replace the front
cover.
7RZDVWH
7RRWKHUPRGXOH