EasyManua.ls Logo

Agilent Technologies 5100

Agilent Technologies 5100
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 133
4 Removal/ Installation, Replacement and Adjustment
Polychromator
NOTE
Periodic cleaning of the Agilent 5110 ICP-OES instruments water- cooling filter is required.
Remove the filter mesh element and clean it by running it under a tap before re-sealing the
system.
CAUTION
The camera assembly is an ESD-sensitive part. ESD handling precautions must be taken
before removing assembly from the instrument.
3 Turn-off the argon gas supply to the instrument.
4 Remove the end plate foam to reveal the end plates.
5 Loosen the 7 pozidrive keyhole screws.
6 Remove the earth lead screw.
7 Remove both end plate covers.
8 Unplug the ribbon cable that is connected from the camera board to the D-range connector
on the camera casting access cover.
9 Unplug the shutter connector.

Table of Contents

Other manuals for Agilent Technologies 5100

Related product manuals