146 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
4 Removal/ Installation, Replacement and Adjustment
Optical alignment
3 Remove the polychromator outer box camera cover and polychromator outer box top cover.
4 Remove the camera rotation access plugs.
5 Remove the rubber plugs from the camera access plate and loosen the four hex head screws
(although maintain enough tension to keep camera hard up against the bracket but still able
to be adjusted.)
6 Go to the Rotation page in the Polychromator alignment software.
7 Open Shutter and ignite the plasma and aspirate blank solution.
8 Under Default set the integration time to 20 msec.
DLA Adjustment ScrewFocus Adjustment
Screw
Wavelength Adjustment
Screw