204 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
7 Service Diagnostics and Fault Finding
ICP Expert Diagnostics
In this example the Poly temp will be logged every 10 seconds (10000ms) if the temperature
falls outside the Range of 34-36°C and then only if the temperature changes by more than
2°C (Delta trigger). In this case the Heater temperature will also be logged for the same Poly
temperature criteria because Heater TempC has been ticked under both “Value triggers” and
also Value “Heater TempC”.