78 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
4 Removal/ Installation, Replacement and Adjustment
Torch loader
5 Remove both gas fittings using 3mm hex key.
6 Remove 4 torx 20 screws with insulators.
7 Remove 2 torx 20 screws and move ignitor holder out of the way.
8 Pull casting forward to release from 2 rear locating dowel pins and pull up and out. (If tight to
remove refer to next step).