General Maintenance 4
5977B Series MSD Troubleshooting and Maintenance Manual 101
To Reconnect the MSD to the 7890 GC
Materials needed
Wrench, open-end, 1/4-inch × 5/16-inch (8710-0510)
Procedure
1 Position the MSD so the end of the GC/MSD interface is near the GC.
2 Reconnect the GC/MSD interface and the LVDS cables (if applicable).
3 Slide the MSD to its regular position next to the GC.
Be careful not to damage the GC/MSD interface as it passes into the GC.
Ensure the end of the GC/MSD interface extends into the GC oven.
4 The foreline pump may be located on the floor, on the lab bench next to or
behind the MSD, or under the analyzer chamber at the back of the MSD.
5 Reinstall the capillary column.
6 Pump down the MSD.
7 Turn on the MS and GC. Enter appropriate temperature setpoints for the
GC/MSD interface and GC oven.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals