EasyManuals Logo

Agilent Technologies 5977B Series Troubleshooting And Maintenance Guide

Agilent Technologies 5977B Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #182 background imageLoading...
Page #182 background image
182 5977B Series MSD Troubleshooting and Maintenance Manual
7 Analyzer
Overview
The analyzer with an XTR source (See Figure 39 on page 184)or the analyzer
with a HES (See Figure 44 on page 193) is the heart of the MSD. It ionizes the
sample, filters the ions, and detects them. The sample components exiting the
GC column flow into the ion source. In the ion source, the sample molecules
are ionized and fragmented. The resulting ions are repelled from the ion
source into the quadrupole mass filter. The mass filter allows selected ions to
pass through the filter and strike the detector. The detector generates a signal
current proportional to the number of ions striking it.
The analyzer is attached to the vacuum side of the side plate. The side plate is
hinged for easy access. The ion source and the mass filter are independently
heated. Each is mounted inside a radiator for correct heat distribution.
Each of the parts of the analyzer is discussed in the following material.
The analyzer has four basic components
The analyzer consists of the following components (See Figure 39 on page 184)
or (See Figure 38 on page 183):
Ion source
Mass filter
Detector
Heaters and radiators

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals